The 2B Technologies POM is a personal ozone monitor that measures ozone by UV absorption (254 nm) over 0–0.5 ppm at 0.1 ppb resolution, with a 3 ppb detection limit and built-in GPS. It is a US EPA Federal Equivalent Method (EQOA-0815-227).

Schedule a callBuild a Configuration

Measures

What the 2B Technologies POM Measures

3 parameters, each linked to its page: the sensors, units, and standards behind it.

Models

2B Technologies POM Models

1 model, with what it measures and the ranges from its datasheet.

ModelDescriptionMeasures & specs
POMUV-absorption ozone monitor (254 nm, folded 'U' optical path) measuring 0–0.5 ppm O₃ at 0.1 ppb resolution, with a 3 ppb detection limit and precision of 1.5 ppb or 2% of reading, whichever is greater. Built-in GPS; serial and USB output at 19200 baud. EPA Federal Equivalent Method.
Range: 0–0.5 ppmResolution: 0.1 ppbMethod: UV absorption at 254 nm (folded 'U' optical path) datasheet

Standards

Standards for What It Measures

Standards that set limits on the parameters it reports, plus any method standard the instrument is built to.

Solutions

Where the 2B Technologies POM Is Used

Solutions whose recommended configuration uses this instrument or whose standards it is built to, followed by solutions whose core panel it partly measures.

On the platform

What the 2B Technologies POM Gets From Aethair

Connected through Thiamis, its readings arrive in the same Environet account as Aethair's own monitors.

Similar

Other Specialty Air Analyzers

Questions

2B Technologies POM FAQ

How does the POM measure ozone?

By UV absorption at 254 nm. A mercury lamp shines through a folded 'U'-shaped optical cell and ozone is quantified from the light it absorbs (Beer-Lambert law). Folding the path lets the handheld reach the same effective path length as 2B Technologies' benchtop 106-L, 202, and 205 ozone monitors.

What is the measurement range and detection limit?

0 to 0.5 ppm ozone at 0.1 ppb resolution. The detection limit is 3.0 ppb in the standard 10-second mode with default adaptive filtering, and precision/accuracy is the greater of 1.5 ppb or 2% of reading. The upper limit was reduced to 0.5 ppm in the current manual revision; older documentation quoting a 10 ppm range is superseded.

Why does the POM also report temperature and pressure?

They are the optical-cell temperature (in Kelvin) and cell pressure (in Torr) measured inside the instrument. The UV-absorption calculation needs both to convert absorbance into an ozone concentration, so the POM streams them alongside every ozone reading. They are internal instrument conditions, not ambient weather values.

Is the POM an EPA-approved method?

Yes. It is designated a US-EPA Federal Equivalent Method for ozone, EQOA-0815-227.

What does the Thiamis integration ingest, and is the GPS position logged?

The driver reads the POM's comma-separated data line over serial at 19200 baud and surfaces ozone, cell temperature, and cell pressure (the first three fields). The built-in GPS logs latitude and longitude with each reading on the device itself, but those coordinates are not ingested as cataloged measurements.